0512-87663997    sales@yosemitech.com

COD

Y552-B Self-cleaning COD Sensor

  • Product Model:Y552-B
  • Time: 2023-05-04 14:54:33
  • Product description:Yoesemite‘s Y552-B self-cleaning COD sensor is based on the UV absorption principle, does not need reagents, will not cause pollution, more environmental protection; Integrated self-cleaning brush, easy to install and use, even long-term online monitoring still has excellent stability.
  • Online Message
  • Email: sales@yosemitech.com

Product Description

Yosemitech Y552-B self-cleaning COD Sensor adopts UV absorption principle, does not need reagents, will not cause pollution, more environmental protection; All-metal structure, compact size. Integrated self-cleaning brush, easy to install and use, even long-term online monitoring still has excellent stability. Chemical Oxygen Demand (COD) Sensor is widely used in surface water monitoring, sewage treatment, seawater monitoring, etc.

Product Features

Digital sensors, digital RS-485 output, Modbus protocol;

Proven UVC LED technology, long lifetime, stable and instant measurement;

Measurement of parameters such as COD, TOC, turbidity and temperature;

With self-cleaning brushes to prevent biological adhesion and longer maintenance intervals.

Product Size

self-cleaning COD sensor size.png

Product Parameters


Name

COD Sensor

Item 

Y552-B

COD Range 

0.5 to 500mg/L equiv.KHP

COD Accuracy

±5% equiv.KHP

COD Resolution

0.01mg/L

TOC Range

0.3 to 200mg/L equiv.KHP

TOC Accuracy

±5% equiv.KHP

TOC Resolution

0.01mg/L

Turbidity Range

0-500 NTU

Turbidity Accuracy

±5% or 0.5NTU

Turbidity Resolution

0.01NTU

Temperature Range

0~50℃

Housing IP Rating

IP68

Maximum pressure 

3 bar

Calibration

One-point or Two-pointsCalibration

Power 

DC 12V -24V,current<50mA (non wiping)

Output 

Support RS-485, Modbus protocol

Size

Φ46mm*176mm

Cable Length

10m (default), customizable

Body materials

316L(custom titanium available)