0512-87663997    sales@yosemitech.com

OIW

Y517-A UV254 OIW Sensor

  • Product Model:Y517-A
  • Time: 2019-09-18 03:56:38
  • Product description:Yoesemitech’s online OIW sensor is based measurement principle of ultraviolet fluorescence method. Compared with other methods, the fluorescence method is proven to be more efficient, fast and repeatable. It is widely used in environmental monitoring, analytical chemistry, limnological and oceanographic biology.
  • Online Message
  • Email: sales@yosemitech.com

Product Description

Yoesemitech’s online OIW sensor adopts ultraviolet fluorescence method measuring principle. Compared with other methods, the fluorescence method is proven to be more efficient, fast and repeatable. It is widely used in environmental monitoring, analytical chemistry, limnological and oceanographic biology.
The sensor has an integrated cleaning brush, which can automatic clean up air bubbles, debris, and biofilm, therefore reduce the interference of contamination on measurement data, make the maintenance period longer, and maintain excellent stability for long-term online use.
Typical applications include: crude and refined oil detection, leakage control at offshore oil pipelines, monitoring of BTXE and polycyclic aromatic compound in fresh water, industrial and municipal sewage works.

Product Features

RS - 485 output, Modbus protocol compatible;

Integrated automatic cleaning brush, eliminating measurement interference ;

Use of unique optical and electronic optical filtering technique eliminates the impact of ambient light on the measurement;

UV LED light source with low power consumption, high stability, low drift and long lifetime;

No affection by particulate matter suspended in water.  

Product Size

OIW Sensor Size.png

Product Parameters


Name

OIW Sensor

Model

Y517-A

Principle

Ultraviolet fluorescence

The laser wavelength

254nm

Fluorescence wavelength

360nm

Range

5ppm(NDSA) equiv. to 1.1ppm(BTEX)

Resolution

0.01ppb

MDL

5ppb(NDSA), 1ppb(BTEX)

Linearity

R²>0.999

Housing IP Rating

IP68

Maximum pressure

3 bar

Temperature Range

0 ~ 50℃

Interface

Support RS-485, Modbus protocols

Sensor Size

Φ 50mm * 176.3mm

Power Requirements

DC 12~24V, current <50mA (when not cleaning)

Installation

Drop-in / flow-through mounting (end NPT3/4 thread)

Cable Length

10m (default), custom length available

Body materials

SS316 (Customizable Ti)

Calibration

zero-point, one-point or two-pointscalibration

Previous:no more Next:Y616-B Oil-in-water Sensor (Crude Oil)